LABORATORY: CNR ISPC – Stone LAB

NAME OF THE INSTRUMENT

High Resolution-Accurate Mass GC-MS and Pyrolysis/ High Resolution-Accurate Mass GC-MS

 

GENERAL DESCRIPTION

Gas chromatograph coupled to a High Resolution-Accurate Mass mass spectrometer (GC-HRA-MS) with Orbitrap technology useful for the analysis of organic materials (e.g. glycerolipids, natural waxes, protein, resinous and polysaccharide material). The combination of the pyrolyzer with GC-MS allows characterizing macromolecules (e.g. natural and synthetic resins) through the recognition of the products obtained from pyrolysis.

The GC-MS analysis involves a different sample preparation (~ 0.1 mg) depending on the materials you want to characterize (lipids, proteins, polysaccharides), which is introduced via an autosampler. Py-GC-MS analysis allows you to analyze the sample as it is or after a derivatization reaction which is introduced via an autosampler into the GC.

 

GC-HRAMS TECHNICAL DETAILS:

  • Mass range: 30 to 3000 m/z
  • Resolution: 100000 at m/z =272
  • Mass accuracy: internal < 1ppm RMS; external <3 ppm RMS
  • Electron ionization (EI) source
  • Chemical ionization (CI) source
  • Scan speed: up to 18 Hz at a resolution setting of 12,500 at m/z 272
  • Sensitivity: ppt

PYROLYZER TECHNICAL DETAILS:

  • Four analytical techniques available:
  1. Evolved Gas Analysis (EGA-MS)
  2. Single-shot analysis
  3. Double-shot analysis
  4. Heart-cut EGA-GC/MS (HC/EGA-GC/MS)
  • Oven temperature control range/temperature stability: ambient temperature +10 to 1050 ºC (1 ºC increments) / within ± 0.1 ºC
  • Heating oven heating rate: maximum 600 ºC/min (every 1 ºC/min)

 

Referent:
Giulia Germinario giulia.germinario@cnr.it

 

Download the full sheet